| 绝缘电阻测量遵循IEC 61557-2标准的测量范围: RISO min= 50kΩ...20 TΩ (I = 1.2 mA, 3 mA or 5 mA) | ||
| 范围 | 分辨率 | 精度 | 
| 0,0...999 kΩ | 1 kΩ | ±(3% 读数+ 10 个字) | 
| 1,00...9,99 MΩ | 0,01 MΩ | |
| 10,0...99,9 MΩ | 0,1 MΩ | |
| 100...999 MΩ | 1 MΩ | |
| 1,00...9,99 GΩ | 0,01 GΩ | |
| 10,0...99,9 GΩ | 0,1 GΩ | |
| 100...999 GΩ | 1 GΩ | ±(3,5% 读数+ 10 个字) | 
| 1,00...9,99 TΩ | 0,01 TΩ | ±(7,5% 读数+ 10 个字) | 
| 10,0...20,0 TΩ | 0,1 TΩ | ±(12,5% 读数+ 10 个字) | 
| 依据测试电压的最大绝缘测量值: | ||
| 电压UISO | 测试 范围 | AutoISO-5000 测试范围 | 
| 50 V | 200 GΩ | 20,0 GΩ | 
| 100 V | 400 GΩ | 40,0 GΩ | 
| 250 V | 1,00 TΩ | 100 GΩ | 
| 500 V | 2,00 TΩ | 200 GΩ | 
| 1000 V | 4,00 TΩ | 400 GΩ | 
| 2500 V | 10,00 TΩ | 400 GΩ | 
| 5000 V | 20,0 TΩ | 400 GΩ | 
| 步进电压电阻测试(SV) | ||
| 电压UISO | MIC-5050 | |
| 50...1000 V | 25 V | |
| 1000...5000 V | 50 V | |
| 导通性(连续性)测试 保护连接和等电位连接阻抗的低压测试 在双向的200mA的电流下对保护导体连接的检测 遵循EN 61557-4标准的测试范围: 0,12...999 Ω | ||
| 范围 | 分辨率 | 精度 | 
| 0,00...19,99 Ω | 0,01 Ω | ±(2%读数+ 3 个字) | 
| 20,0...199,9 Ω | 0,1 Ω | |
| 200...999 Ω | 1 Ω | ±(4%读数+ 3 个字) | 
| 开放终端电压: 4...24V 输出电流在R < 2Ω:Imin > 200mA (ISC: 200...250mA) 测试线电阻归零 双向电流,平均值电阻值已显示 | ||
| 直流(DC)和交流(AC)电压的测量 | ||
| 范围 | 分辨率 | 精度 | 
| 0.0...29,9 V | 0,1 V | ±(2% 读数+ 20 个字) | 
| 30,0...299,9 V | 0,1 V | ±(2%读数+ 6 个字) | 
| 300...750 V | 1 V | ±(2% 读数+ 2 个字) | 
| 频率范围: 45...65Hz | ||
| 电容测量 | ||
| 显示范围 | 分辨率 | 精度 | 
| 1...999 nF | 1 nF | ±(5%读数+ 5 个字) | 
| 1,00...49,99 µF | 0,01 µF | |
| 结果显示顺序:先RISO,后电容 对于电压低于100V的测量,电容测量精度未指定 | ||
| 温度测量 | ||
| 显示范围 | 分辨率 | 精度 | 
| -40,0...99,9 °C | 1 °C | ±(3%读数+ 8 个字) | 
| -40,0...211,8 °F | 1 °F | ±(3%读数+ 16 个字) |